Adaptive averaging for improved SNR in real-time coronary artery MRI

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

3 Author(s)
Sussman, M.S. ; Dept. of Med. Biophys., Univ. of Toronto, Ont., Canada ; Robert, N. ; Wright, G.A.

A technique has been developed for combining a series of low signal-to-noise ratio (SNR) real-time magnetic resonance (MR) images to produce composite images with high SNR and minimal artifact in the presence of motion. The main challenge is identifying a set of real-time images with sufficiently small systematic differences to avoid introducing significant artifact into the composite image. To accomplish this task, one must: 1) identify images identical within the limits of noise; 2) detect systematic errors within such images with sufficient sensitivity. These steps are achieved by evaluating the correlation coefficient (CC) between regions in prospective images and a template containing the anatomy of interest. Images identical within noise are selected by comparing the measured CC values to the theoretical distribution expected due to noise. Sensitivity for systematic error depends on the SNR of the CC(=SNRCCmax), which in turn depends on the noise, and the template size and structure. By varying the template size, SNRCCmax may be altered. Experiments on phantoms and coronary artery images demonstrate that the SNRCCmax necessary to avoid introducing significant artifact varies with the target composite SNR. The future potential of this technique is demonstrated on high-resolution (∼0.9 mm), reduced field-of-view real-time coronary images.

Published in:

Medical Imaging, IEEE Transactions on  (Volume:23 ,  Issue: 8 )