Integrated catheter for 3-D intracardiac echocardiography and ultrasound ablation

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
2 Author(s)
Gentry, K.L. ; Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA ; Smith, S.W.

We present an integrated catheter for 3-D intracardiac echocardiography and ultrasound ablation. The device uses a 9 Fr (3 mm O.D.) catheter lumen with a 14 Fr (4.7 mm O.D.) tip. Real-time 3D imaging is achieved with a forward-looking 112 element, 5 MHz two-dimensional array built on a multi-layer flexible interconnect circuit. Transducer elements typically showed a 21% bandwidth and a pulse-echo insertion loss of 80 dB. The imaging array, attached to our real-time 3-D scanner, was used to image tissue phantoms and a fixed sheep heart. The ablation transducer is a ring of 10 MHz PZT-4 placed around the imaging array. The ring has a 4.5 mm O.D. and a 3.1 mm I.D. for a total area of 33.4 mm2. Using a hydrophone, the spatial-peak, temporal-average intensity of the ablation ring was measured at a maximum of 500 cycles per burst and a PRF of 10.3 kHz to be 16.1 W/cm2. The ablation ring was then used to heat tissue-mimicking rubber with absorption comparable to tissue. A temperature rise of 14°C was recorded. In our final experiment, the imaging array was used to view a 1 cm thick piece of beef muscle with a 12 mm hole. The ablation ring was then used to create a lesion adjacent to the hole.

Published in:

Ultrasonics, 2003 IEEE Symposium on  (Volume:1 )

Date of Conference:

5-8 Oct. 2003