Bayesian reconstructions from emission tomography data using a modified EM algorithm | IEEE Journals & Magazine | IEEE Xplore

Bayesian reconstructions from emission tomography data using a modified EM algorithm


Abstract:

A novel method of reconstruction from single-photon emission computerized tomography data is proposed. This method builds on the expectation-maximization (EM) approach to...Show More

First Page of the Article

Abstract:

A novel method of reconstruction from single-photon emission computerized tomography data is proposed. This method builds on the expectation-maximization (EM) approach to maximum likelihood reconstruction from emission tomography data, but aims instead at maximum posterior probability estimation, which takes account of prior belief about smoothness in the isotope concentration. A novel modification to the EM algorithm yields a practical method. The method is illustrated by an application to data from brain scans.<>
Published in: IEEE Transactions on Medical Imaging ( Volume: 9, Issue: 1, March 1990)
Page(s): 84 - 93
Date of Publication: 31 March 1990

ISSN Information:

PubMed ID: 18222753

First Page of the Article


Contact IEEE to Subscribe

References

References is not available for this document.