Abstract:
Reflective wrist photoplethysmograph (PPG), obtained by a watch or wristband, can provide a natural and unconstrained way for daily life heart rate monitoring. However, r...Show MoreMetadata
Abstract:
Reflective wrist photoplethysmograph (PPG), obtained by a watch or wristband, can provide a natural and unconstrained way for daily life heart rate monitoring. However, reflective wrist PPG often suffers from poor signal quality and various distortions due to daily life motion artifact. In this paper, we analyze the influence of motion artifact on reflective wrist PPG signals, and propose a method to extract reliable heart rate from such distorted PPG signals. The proposed method consists of adaptive filtering, heart rate selection, and motion identification. Experimental results show that our proposed method can generate reliable heart rate values from wrist PPG signals with different types of motion artifact.
Date of Conference: 08-12 June 2015
Date Added to IEEE Xplore: 10 September 2015
ISBN Information: