Abstract:
The loading frame lifting mechanism of the 20 kN dead-weight type force standard machine was improved to enable the one step loading operation at the first force step. Th...Show MoreMetadata
Abstract:
The loading frame lifting mechanism of the 20 kN dead-weight type force standard machine was improved to enable the one step loading operation at the first force step. This modification was achieved by adding a reduction gear to amplify the output torque of the motor driving the loading frame. The one step loading operation allows the evaluation of the time-dependent characteristics of force transducers, and also improves the working efficiency. Some force measuring devices showed a little change of sensitivity upon applying the one step loading operation. It was found that the degree of time dependence in the characteristics of the three force measuring devices examined in this study had only a small impact on the measurement uncertainty.
Published in: 2008 SICE Annual Conference
Date of Conference: 20-22 August 2008
Date Added to IEEE Xplore: 21 October 2008
ISBN Information: